2. 1 Installation Requirements
The following conditions are required for installation of the equipment.
(1) Power source should be as follows (according to your region).
U.S.A. / Europe 220 - 240V (+6% / -10%), 16A, 50/60Hz
For use in Europe: Plug of power cord must meet the standard of each country.
For PERMANENT CONNECTION, a readily accessible disconnect device shall be incorporated in the building installation wiring.
(2) The equipment must be on a dedicated circuit.
The outlet must be near the equipment and easily accessible.
(3) Connect this equipment to a properly grounded outlet.
(4) The site temperature range 10 - 32 degrees C (or 50 - 89 F) with the humidity between
20% - 80% (NON CONDENSING).
Keep the equipment away from water sources, boilers, humidifiers, refrigerators or kerosene (oil) stove.
(5) The installation site must not have open flames, dust or ammonia gases.
(6) The equipment should not be exposed to the direct sunlight.
Please draw curtains to block any sunlight.
(7) Ozone will be generated while this equipment is in use, although the quantity generated is within safe levels. (see certifications) Ventilate the room, if required.
(8) Leveling Bolts on the bottom of the KIP 9900 should touch the floor securely.
The equipment must be leveled. Floor strength must be ample to sustain the weight of the system.
(9) Keep ample room around the equipment to ensure comfortable operation.Required space is noted.